Tips of Using the Seamaty Coagulation Reagent Disc.

 

1.Do the Seamaty reagent discs require warm-up before use?

 

 

Yes.

After taking out the reagent discs from the refrigerator, wait 20 minutes till the disc come back to normal temperature before testing.

The environment of warm-up should be under the normal temperature environment of 10-40 ℃.

Please do not use the hair dryer or other tools to try to warm up the reagent discs in short time.

 

2. Can the seamaty reagent discs be reused?

 

No. The discs are disposable reagent disc. One patient, one disc. The used reagent discs are considered a bio-hazard and should be properly handled as medical waste.

 

3. Can expired discs be used?

 

 

No. Never use a damaged or expired reagent disc.

 

 

4.Add samples into the reagent discs gently, please make sure there is no bubbles in the process.

 

 

5. The samples should be analyzed immediately after collecting the sample.

 

If the samples can’t be tested immediately, the samples should be refrigerated at 2-8°C (36-46° F). And the samples should be analyzed within 2 hours.

 

 

6. After collecting the blood, put the blood sample into the heparin lithium anticoagulant tube immediately then hold the tube upside down 8-10 times, to make sure the sample and anticoagulant mixed together evenly.

Please do not shake the tube intensively.

 

 

7. Please collect blood gently and avoid excessive force. The time of collecting blood should be within 1 minutes.

 

8. The coagulation reagent discs are only suitable for whole blood or plasma samples that are anticoagulated with sodium citrate.

 

9.Avoid hemolysis and coagulation during blood collection, if there is hemolysis and coagulation, the samples should be collected again.

 

 

 

 

 

 

 

 

Pub Time : 2019-09-16 15:55:10 >> News list
Contact Details
Chengdu Seamaty Technology Co.,Ltd

Contact Person: Mr. Bob Zhu

Tel: 86-28-69760308

Fax: 86-28-69760307

Send your inquiry directly to us (0 / 3000)

Request A Quote

E-Mail | Sitemap

Privacy Policy | China Good Quality Veterinary Chemistry Analyzer Supplier. © 2019 - 2022 veterinarychemistryanalyzer.com. All Rights Reserved.